Strainsert
An AS9100C (Including ISO 9001:2008)
Certified Company
Call Now: 610-825-3310
Home
Products
Custom Solutions
Services
Applications
About Us
Catalog & CAD Library
Contact Us
Catalog Quick Links:
Load Pins
Load Cells
Custom Solutions
Force Sensing Bolts/Studs
Force Sensing Tension Links
Instruments
Accessories
Search Catalog
By Keyword
By Item #
All Products
Bi-Axial Load Pins
All Products
>
Load Pins
>
Custom Load Pins
>
Load Pin Optional Features
>
BI-Axial Clevis Pins
Download PDF
Download PDF
Printable Page
Email This Page
Save To Favorites
Please wait…
BI-Axial Clevis Pins
larger image
Bi-Axial Load Pins are designed for applications that measure a radially applied force (Pr) where the load direction is unknown and/or changing in direction. This is facilitated by the installation of 2 perpendicular strain gage bridges (Px and Py), instrumentation, and geometric calculations.
One common use for a Bi-Axial Load Pin is at the hub of a sheave (or pulley) system, which is used to measure line tension. A Bi-Axial Load Pin installation has one end of the line at a fixed angle as a reference while the other end will vary in angle. See Strainsert Load Pin for Sheave Applications Note.
Applications
·
Specifications
·
Crosstalk Description
·
Crosstalk Correction
Applications
Sheave/Pulley Assemblies
Rotating Joints
Hinge Pins
Actuators
Landing Gear
Specifications
Bridge Composition
Full
Bridge Excitation
10 (typical), 12 maximum
Vdc
Bridge Resistance
350 (nominal)
Ohm
Bridge Sensitivity
0.5 and higher
mV/V
Temperature, Zero
±0.01% (nominal) F.S./°F
Temperature, F.S.
±0.01% (nominal) load/°F
Zero Balance
±3.0% (nominal) F.S.
Non-Repeatability
±0.15% FSO (Nominal)
Non-Linearity
±0.50% FSO (Nominal) on Primary axes
Hysteresis
±1 % (nominal) F.S.
Safe Overload
150% (typical) F.S.
Ultimate Strength
300% (typical) F.S.
Minimum Pin Diameter
1 (standard), 0.5 (with limitations) inch
Crosstalk Error, after correction
5% (maximum), <2% (typical) F.S.
Material
Alloy Steels
Higher Strength Stainless
Stainless Steel 17-4, H-1025 (Standard)
Crosstalk Description
Crosstalk error is observed when a radial force (Pr) is applied at varying angles and the sensor output magnitude does not fit the expected circular pattern. Instead, there are higher-than-expected readings at angles between the primary X and Y axes. This error is primarily caused by internal strain gage transverse sensitivity, which is due to physical design and construction constraints. Ideally, the ratio of transverse-to-axial sensitivity is negligible, however, this is not always the case in practice. Crosstalk error is unique and variable to each individual sensor Therefore, crosstalk errors may be marginal and crosstalk corrections may not always be required
Crosstalk Correction
Crosstalk error mitigation efforts during sensor build may be difficult or impossible to achieve physically, regardless of technique or manufacturer quality. Therefore, to minimize the effects of crosstalk error, a mathematical correction is the most complete and preferred method. Strainsert has developed a mathematical crosstalk correction process, which provides realistic data that can be used in force analysis and stress calculations.
The inputs to the crosstalk correction function are the bi-axial strain readings and calibration constants. The calibration constants are determined during factory and/or in-place calibration.
The outputs of the crosstalk correction function are both Px and Py, which may be converted to Pr and angle (θ). This results in the most accurate measurement possible and is in the form of a corrected unit circle (range between 0 and 1 for loads between zero-scale and full-scale, respectively). This output may then be multiplied by the full-scale load value to determine the applied force in load units.
To determine the calibration constants, a factory crosstalk calibration is performed at Strainsert. For each crosstalk calibration, a Strainsert Crosstalk Correction Data report is provided. The crosstalk calibration consists of a series of angle measurements which can be either standard or customized to within a specified quadrant (0° – 90°, 90° – 180°, etc.). If the conditions of the target installation versus factory calibration are larger than expected, additional error may be introduced. Therefore, an in-place calibration is recommended after the clevis pin is installed. Using customer provided data, Strainsert will generate and provide an in-place Strainsert Crosstalk Correction Data report. The inplace calibration constants will then be used instead of the factory calibration constants.
MORE IMAGES
·
sheave-load-pin-variable-angle
DOWNLOADS
Strainsert Bi-Axial Load Pin Application Note
(PDF, 233KB)